Page 67
Setup
5.7 Printer
The printer icon allows you to configure the printer settings for the analyzer. You can
independently configure the printing of analysis results and the frequency of printing for
online TOC and conductivity modes.
Printer connection is made using the 9-pin D-subminiature serial port. The RS-232
interface communicates at 9600 baud, 8 data bits, 1 stop bit, no parity.
The printout report requires a 40-column printer.
• If the printer is purchased from HACH Company, the switches are preset at the
factory.
• If a Seiko Instruments DPU-414 Type II thermal printer is installed, the electronic DIP
switches must be set as listed in Table 13:
Table 13 Printer DIP switch settings
DIP Switch 1
Switch
Setting
Function
1 Off Serial input
2 On High print speed
3 On Auto loading enabled
4 Off Auto linefeed disabled
5 On Enabled setting
6 Off Print density
7 On Print density
8 On Print density
DIP Switch 2
Switch
Setting
Function
1 On 40-column printing
2 On Backup font enabled
3 On Normal characters
4OffZero = Slash
5 On International character set = USA
6 On International character set = USA
7 On International character set = USA
8 Off International character set = USA
DIP Switch 3
Switch
Setting
Function
1 On 8 data bits
2 Off Parity setting = Yes
3 On Parity condition = Odd
4 Off Flow control (X on/X off)
5 Off 9600 baud rate
6 On 9600 baud rate
7 On 9600 baud rate
8 On 9600 baud rate