Page 184
PAT700 Default Settings
Security Settings Default Setting
Enable security Disabled
Enable data logging Enabled
Password expires in # days 60
Remember last # passwords 3
Automatically log off after # minutes 30
Administrator user name Administrator
Administrator ID Admin
Administrator password 123456
User's full name Blank
User ID Blank
User password Blank
Change password Enabled
Bottles Change Settings Default Setting
TOC calibration
Manually accept results Enabled
Return to on-line mode Disabled
Slope limit 10%
R2 limit 0.99
Conductivity calibration
Manually accept results Enabled
Return to on-line mode Disabled
Slope limit from previous 5%
Slope limit from factory calibration 10%
Use calibration resistor Enabled
TOC validation
Return to on-line mode Disabled
System suitability
Return to on-line mode Disabled
Bottle Mode Default Setting
TOC calibration setup Use default settings
TOC calibration custom settings All standards selected, 3 reps
Schedule bottle test Current time