ModBUS Register Information
63
Setup/Analog Output 2 Type 40031
Unsigned
Integer
1R/W
Selects output type (0 = Linear output;
1 = PID control)
Setup/Analog Output 2 Transfer Value 40032 Float 2 R/W Sets the transfer value
Setup/Analog Output 2 Filter 40034
Unsigned
Integer
1R/W
Sets the output filter value in seconds
(0 to 120 sec)
Setup/Analog Output 2
0mA - 4mA
Select
40035
Unsigned
Integer
1R/W
Selects 0mA/4mA for min output
(0 = 0mA; 1 = 4mA)
Setup/Analog Output
2/Linear
Min Setting 40036 Float 2 R/W Sets the min output value
Setup/Analog Output
2/Linear
Max Setting 40038 Float 2 R/W Sets the max output value
Setup/Analog Output
2/PID
Mode 40040
Unsigned
Integer
1 R/W Sets the PID mode (0 = auto; 1 = manual)
Setup/Analog Output
2/PID
Manual Set 40041 Float 2 R/W
Sets the PID manual output value
(0.0 to 100.0%)
Setup/Analog Output
2/PID
Setpoint 40043 Float 2 R/W Sets the PID setpoint
Setup/Analog Output
2/PID
Phase 40045
Unsigned
Integer
1R/W
Sets the PID phase
(0 = Direct; 1 = Reverse)
Setup/Analog Output
2/PID
Proportional
Band
40046 Float 2 R/W Sets the PID proportional band
Setup/Analog Output
2/PID
Integral Time 40048
Unsigned
Integer
1 R/W Sets the PID integral time (min)
Setup/Analog Output
2/PID
Derivative
Time
40049
Unsigned
Integer
1 R/W Sets the PID derivative time (min)
Setup/Relay 1 Source 40050
Unsigned
Integer
1R/W
Selects data source for this relay
(0 = None; 1 = Real Time Clock;
2 = Sensor; 4 = Calculation)
Setup/Relay 1 Sensor Select 40051
Unsigned
Integer
1R/W
Selects Sensor source when
Source = Sensor
(0 = Sensor1; 1 = Sensor2)
Setup/Relay 1
Measurement
Select
40052
Unsigned
Integer
1R/W
Selects measurement on the Sensor
(0 = Meas1 .. 3 = Meas4)
Setup/Relay 1 Type 40053
Unsigned
Integer
1R/W
Selects the relay type
(0=Alarm;1=Control;
2 = Status; 3 = Timer; 4 = Event)
Setup/Relay 1
Transfer
Setting
40054
Unsigned
Integer
1R/W
Selects the transfer value for the relays
(0 = De-energized; 1 = Energized)
Setup/Relay 1/Alarm High Alarm 40055 Float 2 R/W Sets the high alarm setpoint
Setup/Relay 1/Alarm Low Alarm 40057 Float 2 R/W Sets the low alarm setpoint
Setup/Relay 1/Alarm
High
Deadband
40059 Float 2 R/W Sets the high alarm deadband
Setup/Relay 1/Alarm
Low
Deadband
40061 Float 2 R/W Sets the low alarm deadband
Setup/Relay 1/Alarm On Delay 40063
Unsigned
Integer
1 R/W Sets the on delay time
Setup/Relay 1/Alarm Off Delay 40064
Unsigned
Integer
1 R/W Sets the off delay time
Setup/Relay 1/Control Setpoint 40065 Float 2 R/W Sets the controller setpoint
Table 14 Controller ModBUS Registers (continued)