ModBUS Register Information
Setup/Relay 1/Control Phase 40067
Unsigned
Integer
1R/W
Sets the controller phase
(0=Low;1=High)
Setup/Relay 1/Control Deadband 40068 Float 2 R/W Sets the controller deadband
Setup/Relay 1/Control
Overfeed
Timer
40070
Unsigned
Integer
1 R/W Sets the overfeed timer value (mins)
Setup/Relay 1/Control On Delay 40071
Unsigned
Integer
1 R/W Sets the on delay time (sec)
Setup/Relay 1/Control Off Delay 40072
Unsigned
Integer
1 R/W Sets the off delay time (sec)
Setup/Relay 1/Control
Reset
Overfeed
Timer
40073
Unsigned
Integer
1 R/W Resets the overfeed timer
Setup/Relay 1/Event Setpoint 40074 Float 2 R/W Sets the event setpoint
Setup/Relay 1/Event Phase 40076
Unsigned
Integer
1 R/W Sets the event phase (0 = Low; 1 = High)
Setup/Relay 1/Event Deadband 40077 Float 2 R/W Sets the event deadband
Setup/Relay 1/Event On Max Time 40079
Unsigned
Integer
1 R/W Sets the max on time (mins)
Setup/Relay 1/Event On Min Time 40080
Unsigned
Integer
1 R/W Sets the min on time (mins)
Setup/Relay 1/Event Off Max Time 40081
Unsigned
Integer
1 R/W Sets the max off time (mins)
Setup/Relay 1/Event Off Min Time 40082
Unsigned
Integer
1 R/W Sets the min off time (mins)
Setup/Relay 1/Timer Hold Type 40083
Unsigned
Integer
1R/W
Sets which Sensor outputs are affected
during timer on time (0 = None;
2 = Selected Sensor; 13 = All Sensors)
Setup/Relay 1/Timer Sensor Select 40084
Unsigned
Integer
1R/W
Selects which Sensor outputs are being
held/transferred during the timers on time
(this is used when Hold type is set for
single Sensor)
Setup/Relay 1/Timer Hold Mode 40085
Unsigned
Integer
1R/W
Selects hold outputs vs. set transfer value
during timers on time
Setup/Relay 1/Timer Duration Time 40086
Unsigned
Integer
1 R/W Sets the timer on duration time (sec)
Setup/Relay 1/Timer Period Time 40087
Unsigned
Integer
1R/W
Sets the period between timer on events
(mins)
Setup/Relay 1/Timer Off Delay 40088
Unsigned
Integer
1R/W
Sets the time the affected Sensor outputs
are held/transferred after the timer turns
off (sec)
Setup/Relay 1/Status Level 40089
Unsigned
Integer
1R/W
Sets the status level which will trigger the
relay
Setup/Relay 2 Source 40090
Unsigned
Integer
1R/W
Selects data source for this relay
(0 = None; 1 = Real Time Clock;
2 = Sensor; 4 = Calculation)
Setup/Relay 2 Sensor Select 40091
Unsigned
Integer
1R/W
Selects Sensor source when Source
= Sensor (0 = Sensor1; 1 = Sensor2)
Setup/Relay 2
Measurement
Select
40092
Unsigned
Integer
1R/W
Selects measurement on the Sensor
(0 = Meas1 .. 3 = Meas4)
Table 14 Controller ModBUS Registers (continued)