5.2 Show maintenance information of the cleaning module
1. Push menu.
2. Select SENSOR SETUP>[select analyzer]>DIAG/TEST>COUNTERS.
3. Select an option.
Option Description
WIPER REPLACE Shows the remaining number of wiper cycles before a wiper replacement is
necessary.
VIAL TIME Shows the date of the last vial installation or replacement.
Section 6 Maintenance
W A R N I N G
Burn hazard. Obey safe handling protocols during contact with hot liquids.
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the tasks described in this
section of the document.
C A U T I O N
Personal injury hazard. Never remove covers from the instrument. This is a laser-based
instrument and the user risks injury if exposed to the laser.
C A U T I O N
Personal injury hazard. Glass components can break. Handle with care to prevent cuts.
N O T I C E
Do not disassemble the instrument for maintenance. If the internal components must be cleaned or
repaired, contact the manufacturer.
N O T I C E
Stop the sample flow to the instrument and let the instrument become cool before maintenance is
done.
To set the output behavior during maintenance, push menu and select SENSOR
SETUP>TU5x00 sc>DIAG/TEST>MAINTENANCE>OUTPUT MODE.