200 Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA Haemonetics
®
Cell Saver
®
Elite
®
User Manual
273 Partial Bowl Will Be Double Washed
Explanation:
Blood processed using a partial bowl may have a lower hematocrit than blood processed using
a normal full bowl. Because the hematocrit of the bowl contents is lower, there is more
supernatant in the bowl. In order to dilute the larger volume of supernatant, a partial bowl may
require two times the normal saline solution.
Corrective Action:
To change the wash volume, touch Cycle Wash Volume and adjust manually.
274 Partial Bowl Will Be Single Washed
Explanation:
The user has entered Wash based on judgment that the volume of red blood cells in the bowl is
sufficient to warrant a single wash volume.
Corrective Action:
To change the wash volume, touch Cycle Wash Volume and adjust manually.
275 Recovered from System Fault
Explanation:
The system has recovered from a fault.
Corrective Action:
Normal operation can continue. If the problem persists, service is required.
Touch Continue to continue.
277 Protocol Settings Have Been Changed
Explanation:
The selected protocol settings group has been applied. The procedure will continue with the
new settings.
278 Records Exported to USB Flash Drive
Explanation:
The selected procedure record(s) and data log(s) have been exported to the USB flash drive
successfully. The USB device can be safely removed.
Table 16, Event Messages