Chapter 11, Troubleshooting 201
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
279 Procedure Complete
Explanation:
The procedure is complete and the device is ready to power down. The procedure may be
resumed, air may be removed from the RBC bag, and/or records, settings, and help may be
reviewed.
Corrective Action:
To begin a new procedure:
1. If a processing set is loaded, close clamps.
2. Turn power off.
3. Turn power on.
OR
Touch Resume Procedure to resume the procedure.
OR
Touch Remove Air to remove air from the RBC bag.
OR
Access settings, records, or help from the menu.
280 Vacuum Detected in Yellow Line
Explanation:
The manifold pressure sensor detected a high vacuum in the yellow line. The saline bag clamps
may be closed or the saline bag may be empty.
Corrective Action:
1. Ensure the yellow line clamp is open.
2. Check the tubing for kinks, occlusions, and proper placement in the air detector and pump.
3. Replace the saline bag if it is empty.
4. Touch Continue to continue.
Table 16, Event Messages
ID # Event Message Text