204 Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA Haemonetics
®
Cell Saver
®
Elite
®
User Manual
302 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Agitate the blood bag.
3. Touch Continue to fill the bowl.
303 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Touch Continue to collect PPP.
304 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•OPEN the blue PRP line clamp.
•CLOSE the yellow PPP line clamp.
2. Touch Continue to collect PRP.
Table 16, Event Messages