Chapter 11, Troubleshooting 205
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
305 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•OPEN the white air line clamp.
•CLOSE the blue PRP line clamp.
•CLOSE the yellow PPP line clamp.
2. Touch Continue to empty the bowl.
306 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Touch Continue to empty the bowl.
307 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•OPEN the white air line clamp.
•CLOSE the blue PRP line clamp.
•CLOSE the yellow PPP line clamp.
2. Agitate the RBC bag.
3. Touch Continue to concentrate.
Table 16, Event Messages
ID # Event Message Text