206 Chapter 11, Troubleshooting
P/N 120745-IE, Manual revision: AA Haemonetics
®
Cell Saver
®
Elite
®
User Manual
308 Operator Action Required
Explanation:
The bag line clamps must be set to continue Sequestration.
Corrective Action:
1. Set the bag line clamps as follows:
•CLOSE the white air line clamp.
•CLOSE the blue PRP line clamp.
•OPEN the yellow PPP line clamp.
2. Agitate the RBC bag.
3. Touch Continue to concentrate.
309 Blood Bag Is Empty
Explanation:
Air was detected while filling the bowl. Sequestration will continue using fluid from the RBC bag.
Corrective Action:
Touch Continue to continue.
310 Sequestration Cycle Complete
Explanation:
The sequestration cycle has completed.
Corrective Action:
Touch Sequestration to sequester another unit.
OR
Touch Cell Salvage to prepare for Cell Salvage.
OR
Touch End Procedure to end the current procedure.
Table 16, Event Messages
ID # Event Message Text