EasyManua.ls Logo

Haemonetics Cell Saver Elite - Page 207

Haemonetics Cell Saver Elite
220 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 11, Troubleshooting 207
Haemonetics
®
Cell Saver
®
Elite
®
User Manual P/N 120745-IE, Manual revision: AA
311 Operator Action Required
Explanation:
The processing set must be prepared before continuing to Cell Salvage.
Corrective Action:
1. Remove the PPP, PRP, and air bag connection and reconnect the waste bag.
2. Close the blood bag clamps and open the reservoir clamp.
3. Open the saline bag clamp.
4. Touch Cell Salvage to continue to Cell Salvage.
OR
1. Touch Sequestration to return to Sequestration.
312 Confirmation Required
Explanation:
The Cell Salvage menu option was selected. This will stop Sequestration. Confirmation is
required to proceed.
Corrective Action:
Touch Cancel to continue Sequestration.
OR
Touch Continue to continue and prepare for Cell Salvage.
OR
Touch End Procedure to end the current procedure.
400 System Fault
Explanation:
An issue has been detected with the system.
Corrective Action:
Restart the device.
If the problem persists, service is required.
Table 16, Event Messages
ID # Event Message Text

Table of Contents

Other manuals for Haemonetics Cell Saver Elite

Related product manuals