Chapter 2, TEG
®
Analyzer Description 21
Haemonetics
®
TEG
®
6s User Manual P/N 115191-IE, Manual revision: AB
Ethernet port The analyzer may be connected via Ethernet cable to a stand-alone computer
or Ethernet switch or router for service, maintenance and setup. LED lights on
either side of the port have the following functions:
Fuse holder The fuse holder is a cylindrical housing that protects and holds the fuse.
Fuse The 250 Volt 5 Amp fuse is located inside the fuse holder on the back of the
analyzer.
Power jack The power jack is located at the back of the analyzer and is the screw-on
connection point for the power cord that is supplied by Haemonetics. Do not
replace the cord with a substitute. If necessary, contact the local Haemonetics
representative for a replacement. Always ensure that the power cord is
connected to an appropriately grounded power source (100 - 240 volt 50/60
Hz) per your institution’s policy. Use an Uninterruptible Power Supply (UPS)
unit between the analyzer and the power source.
Grounding reliability can be achieved only when the analyzer is connected to a
properly grounded outlet.
On/Off Switch The power switch is located at the back of the analyzer. "I" indicates "on" and
"O" indicates "off".
Left LED - Link Speed Right LED - Link Activity
Amber: operating in 1000 BT mode Blinking green: activity
Green: operating in 100 BT mode Off: no link established
Off: operating in 10 BT mode