Preventative Maintenance
• Preventative maintenance should be completed every six (6) months. The lift should
recommend preventative maintenance if it hasn’t had any preventative maintenance for:
o 1,000 lifts (four or five lifts a day 180 days); or,
o Five (5) hours.
• To recommend preventative maintenance, the Lift will:
o Beep one (1) time every thirty (30) seconds, this can be silenced or over-ridden
by the user by changing a setting, see above.
o Flash “Maintenance” in the first line of the display (regardless of which default
display mode the user has selected).
• To reset the counter which notifies the lift when to signal for preventative maintenance,
see the following section titled, “Resetting the Lift Counter”.
Resetting The Lift Counter
• The lift must be in power off state.
• While pressing both Up/Down buttons on hand control, turn “ON” lift. A minimum of 10
seconds must pass followed by a beep to indicate completion of the reset.
• PM lifts count will be “zero”.
• Use lift as normal.