C800/C1000 – Technical Manual
Rev: 30 May 2024
Preventative Maintenance
Preventative maintenance should be completed every six (6) months. The lift should
recommend preventative maintenance if it hasn’t had any preventative maintenance for:
o 1,000 lifts (four or five lifts a day 180 days); or,
o Five (5) hours.
To recommend preventative maintenance, the Lift will:
o Beep one (1) time every thirty (30) seconds, this can be silenced or over-ridden
by the user by changing a setting, see above.
o Flash “Maintenance” in the first line of the display (regardless of which default
display mode the user has selected).
To reset the counter which notifies the lift when to signal for preventative maintenance,
see the following section titled, “Resetting the Lift Counter”.
Resetting The Lift Counter
The lift must be in power off state.
While pressing both Up/Down buttons on hand control, turn “ON” lift. A minimum of 10
seconds must pass followed by a beep to indicate completion of the reset.
PM lifts count will be “zero”.
Use lift as normal.