Puma 20/40116 | Trouble shooting
Signal Possible cause Remedy
ON/OFF light:
fl ashes 11 times
'STALL' overload fault:
A motor is continually demands too much
power.
Check the drive units.
Note: This fault is often caused by taking obstacles that are
too high, or driving against walls, and door frames, etc.
This fault may also be caused by a diffi cult turn from a
standing position. Check the weight distribution of the
chair.
ON/OFF light:
fl ashes 12 times
System does not 'fi t'.
System modules are not compatible.
Program the entire drive system for the relevant wheelchair
with the aid of the DX-Wizard program on the PC.
Always confi rm the programming by switching the
wheelchair on and off.
Note:
If all the plugs are connected properly and you have used the troubleshooting list but not found the cause, contact the Service
Department of Handicare.
Before changing DX bus cables, fuses and/or modules, remove both fuses from the battery compartment, so that the system is
dead.