32Free Chlorine
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the cuvet.
Powder reagents procedure
Add the content of one packet
of HI 93701 DPD reagent.
Replace the cap and shake
gently for 20 seconds (or 2
minutes in case of seawater
analysis).
Wait for a minute to allow the
undissolved reagent to precipitate
and reinsert the cuvet into the
instrument.
Press READ DIRECT and the display will show “SIP” during
measurement.
The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.