33 Free Chlorine
Liquid reagents procedure
To an empty cuvet add 3
drops of HI 93701A-F DPD1
indicator and 3 drops of
HI 93701B-F DPD1 buffer.
Swirl gently to mix, and
immediately add 10 mL of
unreacted sample. Replace
the cap and shake gently
again.
Reinsert the cuvet into the instrument.
Press READ DIRECT and the display will show “SIP” during
measurement.
The instrument directly displays concentration in mg/L of free
chlorine on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by: Bromine, Iodine, Ozone, Oxidized
forms of Chromium and Manganese.
In case of water with hardness greater than 500 mg/L CaCO
3
, shake
the sample for approximately 2 minutes after adding the powder
reagent.
In case of water with alkalinity greater than 250 mg/L CaCO
3
or
acidity greater than 150 mg/L CaCO
3
, the color of the sample may
develop only partially, or may rapidly fade. To resolve this, neutralize
the sample with diluted HCl or NaOH.
x 3
x 3

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments C 99 series and is the answer not in the manual?

Hanna Instruments C 99 series Specifications

General IconGeneral
BrandHanna Instruments
ModelC 99 series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals