The Instrument displays “SAVING”, stores the calibration values and returns to
measurement mode.
Note: If a new EC calibration is performed, the NaCl calibration is automatically
cleared. A new NaCl calibration is required.
Wrong Standard
If the reading is too far from the expected value, the
“WRONG STANDARD” will scroll. Calibration cannot
be conrmed. In this case check if the calibration
solution has been used correctly or clean the probe
by following the EC PROBE MAINTENANCE (see page
79).
Wrong Standard Temperature
If the temperature is out of the 0.0 to 60.0 ºC range,
“WRONG STANDARD TEMPERATURE” message will
be displayed and the temperature value will blink.
GLP is a set of functions that allows storage and retrieval of data regarding
the maintenance and status of the electrode. The following information can
be viewed on the 3rd LCD line during measurement:
TEMPERATURE SENSOR problem (if there is one),
CAL DUE or CELL FACTOR,
TEMPERATURE COEFFICIENT or T.Ref (User-selected)
TIME
DATE
RANGE and BATTERY
CHARGE STATUS
To view more information, enter GLP mode by pressing GLP key.
EC calibration data is stored automatically after a successful calibration.
Calibration
Messages
EC/TDS
GLP
Information
% NaCl
CALIBRATION