EasyManuals Logo

Hanna Instruments Edge User Manual

Hanna Instruments Edge
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
Maintenance
80
The oxygen probe body is made of PEI.
A temperature sensor provides temperature measurements of the sample.
Use the protective cap when the probe is not in use. To replace the membrane
or rell with electrolyte, proceed as follows:
Remove the protective shipping tube by gently twisting and
pulling it o the body of the probe (see g. 1).
Unscrew the membrane cap by turning it clockwise (see g. 2).
The new membrane cap supplied with the meter should be rinsed
with electrolyte solution. Rell with clean electrolyte solution.
Gently tap the sides of the membrane cap to ensure that no air
bubbles remain trapped. Do not tap the bottom directly with your
nger, as this will damage the membrane.
Make sure that the rubber O-ring sits properly inside the membrane
cap. With the sensor facing down, slowly screw the membrane cap
clockwise. Some electrolyte will overow.
DO Probe
Maintenance

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments Edge and is the answer not in the manual?

Hanna Instruments Edge Specifications

General IconGeneral
BrandHanna Instruments
ModelEdge
CategoryMultimeter
LanguageEnglish

Related product manuals