Maintenance
80
The oxygen probe body is made of PEI.
A temperature sensor provides temperature measurements of the sample.
Use the protective cap when the probe is not in use. To replace the membrane
or rell with electrolyte, proceed as follows:
Remove the protective shipping tube by gently twisting and
pulling it o the body of the probe (see g. 1).
Unscrew the membrane cap by turning it clockwise (see g. 2).
The new membrane cap supplied with the meter should be rinsed
with electrolyte solution. Rell with clean electrolyte solution.
Gently tap the sides of the membrane cap to ensure that no air
bubbles remain trapped. Do not tap the bottom directly with your
nger, as this will damage the membrane.
Make sure that the rubber O-ring sits properly inside the membrane
cap. With the sensor facing down, slowly screw the membrane cap
clockwise. Some electrolyte will overow.
DO Probe
Maintenance