Operational Guide
47
Basic vs
Standard
EC mode
EC Meter
Conguration
Measurement data can be logged using Manual Log on Demand or Manual Log
on Stability or Interval Logging. The measurement logs may be exported to a
thumb drive or PC.
Standard Basic
Measurement Conductivity, TDS, Salinity Conductivity, TDS
Set up Parameters Fully selectable Default values used
Log types
Manual Log on demand
Manual Log on stability
Interval Loggin (Fast,
Medium, Accurate)
Manual Log on demand
Manual Log on stability
EC (Electrolytic Conductivity) meter operation is congured using the SETUP
key with an EC probe connected to the meter. The EC-specic parameters
will be seen inserted into the menu. If Basic mode is congured “On”, the EC
parameter list will be simplied, limiting the parameters the user can change.
See Basic mode for a description of this operation before choosing how to
SETUP the meter.
Parameter Description Choices Default Basic mode
BASIC MODE OFF, ON OFF Available
TEMPERATURE
COMPENSATION
The user
may select
Automatic
Temperature
Compensation
or No TC to
congure
absolute
conductivity
No TC or ATC ATC Not available.
ATC is
automatically
used.
C.F (cm-1) User may enter
actual Cell
factor value (if
known).
0.01 to 9.999
cm-1
1.000 cm-1 Not available.
Automatically
determined
during
calibration.