Specications
85
EC TDS Salinity
Range
0.00 to 29.99 μS/
cm, 3.00 to 29.99
μS/cm, 300. to
2999. μS/cm,
3.00 to 29.99
mS/cm, 30.0 to
200.0 mS/cm,
up to 500.0 mS/
cm, absolute *
conductivity
0.00 to 14.99 ppm
(mg/L), 15.0 to 149.9
ppm (mg/L), 150. to
1499. ppm (mg/L), 1.50
to 14.99 g/L, 15.0 to
100.0 g/L, up to 400.0
g/L absolute * TDS (with
0.80 factor)
0.0 to 400.0% NaCl
**, 2.00 to 42.00
PSU **, 0.0 to 80.0
g/L **
Resolution
0.01 μS/cm, 00.1
μS/cm, 0.01 μS/
cm, 0.01 mS/cm,
00.1 mS/cm
0.01 ppm, 0.1 ppm, 1
ppm, 0.01 g/L, 0.1 g/L
0.1% NaCl, 0.01
PSU, 0.01 g/L
Accuracy
@ 25 °C / 77 °F
±1% of reading
(±0.05 μS/cm or 1
digit, whichever is
greater)
±1% of reading
(±0.03 ppm or 1 digit,
whichever is greater)
±1% of reading
EC Calibration
1 Cell Factor calibration;6 standards available: 84, 1413 μS/cm, 5.00,
12.88, 80.0, 111.8 mS/cm, 1 point offset: 0.00 μS/cm
Salinity
Calibration
%NaCl (only)
Single point with HI 7037L Standard
Temperature
compensation
Automatic
NoTC - can be selected to measure
absolute conductivity.
Specications