Channel  40
Prop. control. Editable default and boundary values.
Control parameters Measured parameter Default Minimum Maximum *
Dead Band
pH 0.20 pH 0.00 pH
5 % of measured
range but no more
than Deviation value
divided by 5
ORP 10 mV 0 mV
EC 400.0 μS 0.000 μS
DO 20 %Sat 0.0 %Sat
* Maximum limit can be different from given values depending on probe measurement range (0.5 %)
Setup for PID control
• Press Setup for the options submenu to be displayed.
Press the
ud keys to move between Mode, Deviation, Control Period, Reset Time, Rate Time, Dead
Band, and Dead Band Gain.
Mode: press the virtual keys, to choose PID Low or PID High.
Press the d key, to select Deviation.
With Deviation highlighted, press Set. The present value will blink permitting editing using the ud
keys.
• Press CFM, to save.
Press the
d key, to select Control Period.
With Control Period highlighted, press Set. The present value will blink permitting editing using the
ud keys.
• Press CFM to save.
Press the
d key to select Reset Time.