41  Channel
With Reset Time highlighted, press Set. The present value will blink permitting editing, within the boundary
values, using the ud keys. The default value disables the Integrative contribution.
Press the
d key to select Rate Time.
With Rate Time highlighted, press Set. The present value will blink permitting editing, within the boundary
values, using the ud keys. The default value disables the Derivative contribution.
• Press CFM to save.
Press the
d key to select Dead Band.
With Dead Band highlighted press Set. The present value will blink permitting editing, within the boundary
values, using the ud keys.
• Press CFM to save.
Press the
d key, to select Dead Band Gain.
With Dead Band Gain highlighted, press Set. The present value will blink permitting editing, within the
boundary values, using the ud keys.
• Press CFM to save.
PID control. Editable default and boundary values.
Control parameters Measured parameter Default Minimum Maximum
Deviation
pH 1.00 pH 0.02 pH 6 pH
ORP 50 mV 2 mV 2000 mV
EC 2.000 mS 0.002 μS 500.0 mS
DO 100.0 %Sat 0.2 %Sat 300.0 %Sat
Control Period 1.00 minute 10 seconds 30.00 minutes
Reset Time 16:40 hours 10 seconds 16:40 hours