Cleaning  58
Int. Trigger
Option: Disabled, Timer, Schedule
With Int. Trigger selected, press Modify for the drop‑down options list to be displayed.
Use the
ud keys to scroll between options.
• Press Select to save option.
When set on Timer, cleaning cycle will proceed following the time period set in the parameter Cleaning Interval.
Schedule
If Int. Trigger is selected, options are Disabled or Timer, NA will be seen.
If Int. Trigger is set to Schedule, options are On or Off.
With Schedule On selected, press Setup to configure a cleaning schedule.
Set up to three start times per day for the cleaning cycle to start.
Enable the days of the week for the cleaning cycle to be done.
Press the
b key to save and exit schedule.
Cleaning Interval
Option: 1 to 1440 min. (as 1 to 59 min. and 1:00 to 24:00 h), if Timer is selected as an Int.Trigger
NA, if Schedule is selected as Int. Trigger
With Cleaning Interval selected, press Set, to modify.
Use the
ud keys to modify the flashing digit.
• Press CFM to save.