Introduction 6
Main Features
Process input
Smart probes with RS‑485 connection. Automatic probe recognition and upload of
configuration calibration, and measurement data.
User interface Easily navigable main menu and submenus
Analog Outputs
(programmable)
Two or four, depending on the controller model
` 0 20 mA
` 4 20 mA
Alarm Relay Activates on errors and programmable alarm conditions
Control relays
(programmable)
Three (two SPDT + one SPST) or five (two SPDT + three SPST) contact outputs,
depending on the controller model
` 5 A 250 Vac
` 5 A 30 Vdc
Cleaning function
Integral water or chemical cleaner control
Configurable simple or advanced cleaning:
` triggered manually
` triggered at a set time interval
` scheduled for a specific day of the week
` triggered by a digital input
Blowers, water jets, washers (user supplied)
LCD display
Backlit dot matrix
With virtual key function
Enclosure Rugged molded housing with hinged front panel
Additional Features
Hold mode
Automatic mode for entering calibration, Setup, and cleaning cycle
Manually triggered or triggered via an external digital input
h key
User help key, opens a guide for diagnosing a problem or troubleshooting
Security access code Protected calibration and setup settings
Languages
English (default), Deutsch, Español, Français, Italiano, Magyar, Nederlands,
Português
Remote control Monitor and control of process parameters through RS‑485 (Modbus RTU)
USB‑C port USB for exporting (or importing) data with a flash drive and for software update
Manual mode
Used to exercise relays and analog outputs
Useful for:
` setting up the system
` priming a pump
` checking wiring
` during general maintenance
Default option when the industrial application requires manual input
As a safety feature, a 60‑minutes timeout to exit manual mode is implemented (relays
and analog outputs return to their previous value)