7 Introduction
Data & Event logger
The controller automatically logs the process control information in an interval
log, and various event alarms and errors in an event log
Logged data can be retrieved and events visualized on the screen, in Log Recall menu
Interval logs store up to 8600 records per lot, maximum lot number is 100 lots
Logging interval can be set in the General settings menu
Logged data includes:
` measurement variables and temperature measurements
` last calibration data
` setup configuration
` start/end date and time
` previous values
` event data and event code
Event log can store up to 100 records of events, alarms, errors‑related data
Log files can be uploaded to a USB flash drive via USB‑C port
Calibration
pH
Up to three‑points standard calibration with selection from two buffer groups:
` Hanna Instruments: 1.68, 4.01, 7.01, 10.01, 12.45 pH
` NIST: 1.68, 4.01, 6.86, 9.18, 12.45 pH
Conductivity
Up to two‑points user calibration with selectable calibration points:
` 0.000 μS/cm for offset
` 84.0 μS/cm, 1413 μS/cm, 5.00 mS/cm,12.88 mS/cm for 0.1/cm cell
` 80.0 mS/cm, 111.8 mS/cm additional standards for 1.0/cm cell
DO (Dissolved Oxygen)
Up to two‑points standard calibration
Single point process calibration for all supported parameters
Last calibration data stored on the probe and can be visualized in the Cal Data window
Calibration reminder can be scheduled (1 to 99 days) or set as Off
Safety
• When in Setup or Calibration mode, without making any changes or pressing any
key, the controller will return to Measure mode and restart control.