EasyManua.ls Logo

Hardy Instruments A Series - Cal Low Command (CALLOWCMD); Cal High Command (CALHIGHCMD); C2 Cal Command (C2 CALCMD)

Hardy Instruments A Series
88 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HI 1756-WS WEIGH SCALE MODULE
4-14
Channel Number: 0 or 1
Number of Elements: 8
Error Return Values: None
Cal Low Command
(CALLOWCMD)
The Cal Low Command sets the “calLowCount”
parameter to the current A/D average counts when
doing a hard calibration. An Integrated Technician
function gets called during low calibration.
Command Number: 0x64 (Hexadecimal)
Channel Number: 0 or 1
Number of Elements: 8
Error Return Values:
STATUSWORD - there was a conversion error,
weight in motion or an A/D error or all three.
HARDCALFAILCOUNTS - less than 100
counts between the zero and the span weights.
Cal High
Command
(CALHIGHCMD)
The Cal High Command - Sets the Span Weight
parameter to the current A/D average counts when
doing Hard CAL.
Command Number: 0x65 (Hexadecimal)
Channel Number: 0 or 1
Number of Elements: 8
Error Return Values:
STATUSWORD - there was a conversion error,
weight in motion or an A/D error or all three.
HARDCALFAILCOUNTS - there are less than
100 counts between the zero and the span
weights.
C2 Cal Command
(C2CALCMD)
The C2 Cal Command - Performs a C2 Calibration.
Uses CalLowWeight as the reference point.
Command Number: 0x66 (Hexadecimal)
Channel Number: 0 or 1
Number of Elements: 8