Chapter 4 - Setup
4-23
Setting the Span
Weight Value
The Span Weight is a reference point derived from an
actual measured weight. This should not be confused
with the Scale Capacity. If you have a 100 pound
weight and you place it on the scale, the Span Weight
would be 100 pounds.
Setting the
WAVERSAVER
®
Value
There are 5 selectable levels. 0 provides the least
vibration immunity with the fastest response time. 4
provides the most vibration immunity with the slow
-
est response time. Default setting is 2.
Immunity Setting
7.5 Hz 0
3.5 Hz 1
1.0 Hz 2
0.5 Hz 3
0.25 Hz 4