4.4 Connecting a transducer
Connecting to an MP60: HBM torque transducers T10F-SF1, T10F-SU2,
T4WAS3, incremental transducer, frequency
generator
Connect to an MP07: HBM-torque transducer T10F-KF1, T30FNA,
T32FNA, T34FN, T36FN
Torque measurement (symmetric signals)
(torque flange T10F-SF1, T10F-SU2)
Supply voltage (18 V - 30 V)
Measurement signal for torque, frequency output F1 (+)
Calibration signal trigger (approx. 5 V)
Ground
Plug 1
Md
Cable shielding
Supply voltage (0 V)
Measurement signal for torque, frequency output F1 (-)
23
7/5
6
4
1
BK
BU
WH
RD
GN
GY
YE
5
6
13
12
Hsng.
14
Measurement signal for rotation speed, 0° , F1 (+)
Ground
Plug 2
n
Cable shielding
81
3
6
7
BK and BU
RD
GN
WH
GY
Measurement signal for rotation speed, 0° , F1 (-)
Measurement signal for rotation speed, 90° , F2 (-)
Measurement signal for rotation speed, 90° , F2 (+)
Rotation speed / angle of rotation measurement (symmetric signals);
(Torque flange T10F-SF1, T10F-SU2)
8
12
14
13
Hsng.
MP60
MP60
8
15
Torque
Rotation speed / angle of rotation
Connection A Connection B
Connection A
Connection B
CAUTION
The 24V external supply voltage
must also be connected to terminal plug 4
(power supply for T10F-SF1, T10F-SU2)
Terminal plug 4
Transducer
Transducer error
7
Transducer error
7
Fig. 4.4: Connection for MP60