General Notice for FU / CCI - ERROR 61 to FU / CCI - ERROR 69
Error
consequence
Frequency converter switches independently.
The rotor freewheels, coasting.
No further user operation is possible.
Error code reset
Mains switch is OFF.
Switch mains switch to ON after 1 min.
Measurement
Also see at frequency converter -X501 (PIN 4) and
control panel –X1 (PIN 13).
FU / CCI - ERROR 61
Error Error in computing section.
Cause of error
Flat ribbon cable is defective.
Frequency converter is defective.
FU / CCI - ERROR 62
Error Undervoltage. Mains voltage less than 20% as nominal
voltage.
Cause of error
Supply voltage too low, see chapter "Short the mains choke
coil (L1)".
Flat ribbon cable is defective.
Frequency converter is defective.
Measurement Also see at frequency converter, U
DC
.
FU / CCI - ERROR 63
Error Overcurrent.
Cause of error
Short circuit in motor.
Motor impedance is too low.
Flat ribbon cable is defective.
Frequency converter is defective.

Table of Contents

Other manuals for Hettich EBA 21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich EBA 21 and is the answer not in the manual?

Hettich EBA 21 Specifications

General IconGeneral
BrandHettich
ModelEBA 21
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals