19/42
FU / CCI - ERROR 64
Error Voltage in intermediate circuit:
>410 V DC at 230 V
>205 V DC at 115 V
This error normally only occurs when the drive is being braked.
Cause of error
Brake resistor is defective.
Flat ribbon cable is defective.
Frequency converter is defective.
Measurement Also see at frequency converter, U
DC.
FU / CCI - ERROR 67
Error Only centrifuges with 115 V.
Overtemperature in the motor. The cable “overtemperature” in
the motor has high impedance.
Cause of error
Overtemperature switch opens because of overtemperature
in the motor
Flat ribbon cable is defective.
Frequency converter is defective.
Motor is defective
Measurement Also see at frequency converter, remove plug S2 and measure
at the plug
Switch closed: 0
opened:
FU / CCI - ERROR 68
Error Overtemperature in frequency converter.
Cause of error
Insufficient heat abduction from frequency converter to
centrifuge housing. There is no, or not enough, heat-
conducting paste between frequency converter and
housing.
Full-load operation and an ambient temperature > 45°C.
Flat ribbon cable is defective.
Frequency converter is defective.
FU / CCI - ERROR 69
Error EEPROM in frequency converter is defective.
Cause of error
Flat ribbon cable is defective.
Frequency converter is defective

Table of Contents

Other manuals for Hettich EBA 21

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich EBA 21 and is the answer not in the manual?

Hettich EBA 21 Specifications

General IconGeneral
BrandHettich
ModelEBA 21
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals