FU / CCI - ERROR 84
Error Frequency converter recognises excess speed.
During rotation the speedometer pulses (6 per revolution) are
controlled by the frequency converter. This control is independ-
ent from the control panel.
The frequency converter switches the centrifuge off, when the
maximum speed of the rotor is exceeded more than 500 rpm.
Cause of error
Flat ribbon cable is defective.
Frequency converter is defective.
Measurement Also see at supply board-X4 (PIN 4) and frequency converter -
S501 (PIN 8).
FU / CCI - ERROR 85
Error “Watchdog” in frequency converter
Discrepancy in program procedure
Cause of error
Flat ribbon cable is defective.
Frequency converter is defective.
Error code reset Perform a MAINS RESET.