CP : control panel, FC : frequency converter, SB : supply board, CB : cooling board, CC : control cable, LL : lid locking, BC : braking chopper,
BR : brake resistor, MR : mains reset
21/49
FU / CCI - ERROR 69
Error cause EEPROM in FC is defective.
Error remedy
CC is defective.
FC is defective
FU / CCI - ERROR 84
Error cause FC recognises excess speed.
During rotation the speedometer pulses (6 per revolution) are
controlled by the FC.
The FC switches the centrifuge off, when the maximum speed
of the rotor is exceeded more than 500 rpm.
Error remedy
CC is defective.
FC is defective.
Measurement Also see at SB-X4 and FC S501 (PIN 8).
Error code reset Perform an MR.
FU / CCI - ERROR 85
Error cause “Watchdog” in FC
Discrepancy in program procedure
Error remedy
CC is defective.
FC is defective.
Error code reset Perform an MR.