EasyManuals Logo

Hettich UNIVERSAL 320 R User Manual

Hettich UNIVERSAL 320 R
81 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
AB1401en_SARev.: 04 / 11.2023
10 / 44
Device overview
Ambient conditions (EN / IEC 61010-1):
Installation site indoors only
Altitude up to 2000 m above sea level
Ambient temperature 2°C to 35 °C
Humidity maximum relative humidity 80%for temperatures up to 31°C,
decreasing linearly to 50%relative humidity at 40°C.
Overvoltage category
(IEC 60364-4-443)
II
Pollution level 2
Device protection class I
not suitable for use in potentially explosive atmospheres.
EMC:
Emitted EM interference,
EM interference immunity
EN/IEC61326-1
Class B
FCC Class B
Noise level
(rotor-dependent)
£68dB(A)
Dimensions:
Width 401 mm
Depth 529 mm
Altitude 346 mm
Weight approx.31kg
Manufacturer Andreas Hettich GmbH & Co.KG,
D-78532 Tuttlingen
Model UNIVERSAL 320R
Type 1406 1406-01
Mains voltage (±10%) 200-240V1 240V1 115-127V1
Mains frequency 50 Hz 60 Hz 60 Hz
power consumption 800 VA 950 VA
Power consumption 4.0A 8.0A
Refrigerant R452A
max. capacity 4 x 200 ml
max. permissible density 1.2kg/dm³

Table of Contents

Other manuals for Hettich UNIVERSAL 320 R

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hettich UNIVERSAL 320 R and is the answer not in the manual?

Hettich UNIVERSAL 320 R Specifications

General IconGeneral
BrandHettich
ModelUNIVERSAL 320 R
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals