max. speed (RPM) 16000
max. acceleration (RCF) 24900
max. kinetic energy 9800 Nm
Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
No
Ambient conditions (EN / IEC 61010-1):
Installation site indoors only
Altitude up to 2000 m above sea level
Ambient temperature 5°C to 35 °C
Humidity maximum relative humidity 80%for temperatures up to 31°C,
decreasing linearly to 50%relative humidity at 40°C.
Overvoltage category
(IEC 60364-4-443)
II
Pollution level 2
Device protection class I
not suitable for use in potentially explosive atmospheres.
EMC:
Emitted EM interference,
EM interference immunity
EN/IEC61326-1
Class B
FCC Class B
Noise level
(rotor-dependent)
£64dB(A)
Dimensions:
Width 407 mm
Depth 698 mm
Altitude 346 mm
Weight approx.52kg