213960EN(1) - P290 Mattress Instructions for Use Page 9
Destination
Specifications
Front and Back Views of the P290 Control Unit
The control unit may generate noise and vibration that can cause discomfort to the
patient.
Rated power
0.07 A @ 220-240 Vac 0.17 A @ 120 Vac
Maximum energy
consumed
5.2 Wh
Operation of the device Continuous
Sleeve material ABS PC
Device: sound pressure level
/ sound power (ISO 3744)
45 dB(A) / 61 dB(A)
Fuses T1AL, 250V
Compression 0-10,5 kPa
Compressor flow rate 10 l/min
IEC 60601-1 Classification Classe II
Degree of protection
provided by the cover
(IEC 60529)
IP41: Protected against access to hazardous parts with a
wire and against vertically falling water drops
Protection against
inflammable anesthetic
mixtures
Not for use with flammable anesthetics.
Characteristics Values
Control Panel
Hose Connector
Fuses
Filter
Main Switch
Power Socket
Handle Attachment