Page 8 213960EN(1) - P290 Mattress Instructions for Use
Technical Specifications
Specifications are subject to change.
Essential performances of the device
The P290 is a therapeutic mattress. It has two operating modes: continuous low
pressure (CLP) and alternating low pressure (ALP). The mattress firmness is adjustable
from the control unit.
Mattress Topper and Mattress
Characteristics Values
Mattress Topper and Mattress
Mattres model
P290A1 P290A2
P290A3
Length (inflated) 200 cm / ~ 80’
Width (inflated) 90 cm / ~ 36’
Height (inflated) 10 cm / 4" 17 cm / 7"
Weight 6,1 kg / 13,4 lb 12 kg / 26,5 lb 7,5 kg / 16,5 Ib
Range in transport mode 2 hours
Upper cover (applied part of
the device)
Polyurethane coating on polyester material, low-friction,
stretchable in both directions, breathing, bacteriostatic,
fungistatic and antimicrobial; can be wiped and washed.
Bladders Polyurethane
Safe Working Load is the
greatest permissible load
that can be applied after
which a damage may occur
on the mattress.
Safe working load,
including the total weight
of the patient, accessories
and the load supported by
these accessories .
Pressure relief valve
threshold
1.5 psi / 10,3 kPa
Degree of protection
against electric shock
Type BF applied part
Cushions pressure.
Adjustable range
from 8"H
2
O to 30"H
2
O
Control Unit
Control Unit model
212290 212291
Dimensions 29 x 20 x 12 cm / 11.4" x 8" x 4.6"
Weight 2.2 kg / 5 lb
Power supply 220-240 V~, 50/60 Hz 120 V~, 50/60 Hz