213960EN(1) - P290 Mattress Instructions for Use Page 31
Moving the Device
Moving the Device
Moving the Patient on the Bed in Transport Mode
1. Press the button to select Continuous low pressure mode.
The indicator light of the
symbol comes on.
2. Wait until the pressure is perfectly stable and the compressor stops. This must not
take any longer than 5 minutes.
These two steps guarantee that the system deflates in a uniform manner.
3. Switch off the control unit by pressing .
The indicator light of the
symbol
goes out.The Mains power fault audible
indicator sounds and the indicator light comes on.
The others indicator lights are off.
If an electronic module is disconnected when the device is switched on, the module
may be destroyed.
4. Deactivate the Mains power fault audible indicator by pressing (See “Audible
Indicator Silence” page 28).
5. Disconnect the device from the wall outlet.
The device automatically switches to transport mode and remains
inflated.
Never pull on the power cord or you may damage it. A damaged
power cord is an electric shock hazard.
Make sure that the cord does not drag along the floor, do not run over
it when moving the bed frame and take care not to trip over it. If
necessary, use the attachments provided with the device.
6. Safely stow the power cord.