213960EN(1) - P290 Mattress Instructions for Use Page 39
Disinfecting
Servicing
To avoid damage due to high pressure or an abnormal surface temperature, take the
following precautions:
• Use low pressure steam on the electric parts.
• Do not use accessories such as high pressure
hoses (A). It is preferable to use soft non-
metallic brushes (B) and wipers (C) in such a
way as to reduce the pressure to an acceptable
level.
• Prevent water and steam from getting into
connectors that are not in use.
• Do not brush and apply reduced pressure to
labels and markings.
• Dry carefully and look for signs of water ingress.
• Test the device before reusing it.
Do not steam clean the inside parts but only the outside parts.
Do not wash the zip fasteners with steam.
The upper cover can be machine-washed, but it must not always be machine-
washed, as doing so will reduce the life time of the components. The upper cover
should only be machine-washed if particular risks of infection occur. If this method
is used, choose the wash and spin cycle with reduced mechanical action.
Proceed with the following final steps before reusing the cleaned and disinfected
device.
Performing the Final Steps
•Always remove all traces of the products used when washing or
disinfecting the device.
•Make sure that all the parts of the device are perfectly dry before
installing to avoid any risk of condensation forming inside.
Maximum temperature: 70°C
Delicates
Wash by hand
Only clean with oxygenated agents
Do not iron
Do not dry clean
Can be tumble dried
Moderate temperature: 50°C
Cannot be tumble-dried