11
―――――――――――――――――――――――――――
Chapter 3 Measurement Procedures
――――――――――――――――――――――――
WARNING
Before using the product check that the body of
the product is not damaged. Also make sure that
the insulation on the test leads is undamaged and
that no bare conductors are improperly exposed.
Using the product in such conditions could cause
an electric shock accident may occur, so contact
your dealer or Hioki representative for repair.
NOTE
3.1 Pre-Operation inspection
To avoid the possibility of electric shock or incorrect
measurement, check the following items before using
the product. If the operation check reveals any
abnormalities, stop the check immediately and do not
use the product.
For voltage measurement, short the test leads and
check that 0 V is displayed.
For Measuring Resistance or Continuity Check,
short the test leads and check that 0 Ω is displayed.
Measure a test item with a known value (battery,
AC supply, resistor, etc.) to confirm that the
known value can be displayed.
Periodic calibration and inspecton is necessary in
order to ensure that this product operates according
to its product specifications.