210 Cannot Save (File Lock) Do not write, edit, or delete a file in the PC card
using the FTP during automatic text save or at
the start of the real-time save operation.
300 Cannot START. Measurement cannot be started from a screen
displayed by pressing the
CARD button.
360 Interval has been
changed. (humid)
Humidity measurement cannot be performed un-
less the recording interval is from 5 sec. to 1 hr.
Therefore, the recording interval is automatically
adjusted.
361 Cannot be faster than 5s.
(humid)
When humidity is measured on any channel, a
recording interval less than 5 sec. cannot be set.
362 Measurement range was
changed.
When upper and lower limits have been set, the
range is automatically changed to the optimum
range that corresponds with the upper and lower
limits.
364 Saving was interrupted. Operation was forcibly interrupted while saving
text.
370 Cannot change while
measuring.
Press the
STOP button twice to stop measure-
ment, then change the setting.
375 Invalid event marker. You have attempted to go to an event marker
that is no longer in memory.
382 No waveform data. There is no waveform data to display on the
Waveform Screen. Data must be acquired be-
fore processing can start.
396 Out of range. (variable) The settable area for upper and lower limits has
been exceeded. Enter an appropriate value for
the upper or lower limit. (variable)
397 Out of range. (scaling) Input the appropriate signal level. (scaling)
398 A-B cursor positions in-
valid.
Move the cursors to valid positions.
520 Bad MAC address. The MAC address is illegally rewritten. Contact
us.
521 Bad IP address. Check the IP address.
522 Bad server IP address. Check the server's IP settings.
523 Can not connect to server. Check the settings and connection.
524 Can not connect to 9334. Check the settings and connection of the 9334
LOGGERCOMMUNICATOR.
525 Connection timed out. Check the connected device.
526 Transfer was aborted. Check the connected device.
527 Network error. Check the instrument and connected device.
528 Server not found or DNS
failed.
Check the DNS IP address or the line connec-
tion.
529 DHCP failed. Check the connected device.
530 Password error. Check the password. Password recognition is
case-sensitive.
531 Can not change while
server is working.
Stop the monitor server and edit the setting.