3.8 Using D/A Output
95
The instrument provides three analog output terminals (D/A output) for which the output parameter can be
selected.
See: "Appendix 2 Detailed Specifications of Output" (p. A5)
Selectable output parameters and output voltages
You can select three of the following parameters. (Any channel or sum can be selected for each.)
Setting D/A output parameters
3.8.2 D/A Output
Voltage, current, active power, apparent power, reactive power, time
average current, time average active power
±2 V DC for ±100% of the range
Power factor ±2 V DC for ±0.0000 or 0 V DC for ±1.0000
Phase angle, inter-channel voltage fundamental wave phase differ-
ence, inter-channel current fundamental wave phase difference
0 V DC for 0.00°, ±2 V DC for ±180.00°
Voltage ripple rate, current ripple rate, total harmonic voltage distor-
tion, total harmonic current distortion
+2 V DC at 100.00%
Voltage crest factor, current crest factor +2 V DC at 10.000
Frequency (varies with measured value) +2 V DC per 100 Hz from 0.1000 Hz to 300.00 Hz
+2 V DC per 10 kHz from 300.01 Hz to 30.000 kHz
+2 V DC per 100 kHz from 30.001 kHz to 220.00 kHz
Efficiency +2 V DC at 200.00%
Current integration, active power integration ±5 V DC at (range) × (integration set time)
1 Press .
2 Press to display the settings
screen shown to the left.
3 Press or to select the rectifier,
channel, parameter, and display area.