3.11 When PEAK OVER, o.r, or the Unit Indicator Flashes
116
3.11.3When the Unit Indicator Flashes
Status Solution
The unit indicator flashes during
averaging processing
The average value being displayed contains [o.r] data. If no [o.r] data is present during
averaging processing, the unit indicator will not flash. When [o.r] is displayed, internal
data for which the voltage or current exceeded 140% of the range, or the active power
exceeded 196% of the range, will be used as-is in calculating the average value.
The integrated value or time aver-
age value unit indicator flashes
Reset integrated values, change the range, and repeat integration. If no PEAK OVER
condition occurs during integration, the indicator will not flash.
See: "3.3 Integration" (p.62)