4 Press or to set the integra-
tion time.
Setting range: 0 to 9,999.59
(When the setting display indicates 0000.00, integra-
tion will be performed until 10,000 hours has
elapsed.)
This indicates that the integration time cannot be changed. Take the following action to address the issue:
Status Solution and reference for more information
During integration operation
(RUN lamp lit up or flashing)
The integration time cannot be changed until the integrated value is reset (so that
the RUN lamp turns off).
See: Canceling integration (resetting integrated values) (p.65)
During display hold operation or
maximum value/minimum value
hold operation (HOLD, MIN, or
MAX lamp lit up)
The integration time cannot be changed until this operation is canceled (so that the
HOLD lamp turns off).
See: Canceling the display hold state (p.108)
Returning to the instantaneous value display (p.109)
If the integration time does not flash