6-15 Stream alignment
Adjust the horizontal direction and vertical direction positions.
(a) Horizontal direction adjustment procedure
Slightly loosen the horizontal direction lock screws (2).
As to the screw loosening, please see the precautions below.
Turn the horizontal direction adjustment screw and adjust the position of the makeup.
When you want to move in the minus electrode direction : Turn clockwise
When you want to move in the plus electrode direction : Turn counterclockwise
Adjust so that the stream is approximately at the center of the gutter.
After adjustment, tighten the horizontal direction lock screws (2).
Horizontal direction
lock screw
Horizontal direction
adjustment screw
Plus deflection
electrode
Minus deflection
electrode
Stream
Gutter
Horizontal
direction
Vertical direction
adjustment screw
Vertical direction
lock screw
Stream
Gutter
Vertical
direction
3
2
1
3
(b) Vertical direction adjustment procedure
Slightly loosen the vertical direction lock screw (2).
Turn the vertical direction adjustment screw and adjust the position of the stream.
When you want to move to the bottom of the gutter : Turn counterclockwise
When you want to move to the top of the gutter : Turn clockwise
Adjust so that the stream is approximately at the center of the gutter.
After adjustment, tighten the vertical direction lock screw (2).
2
1
3
At the end of adjustment, press Abort .
4
PERCAUTIONS
Horizontal direction lock screw.
Indication of the
loosening
[60 degree to
120 degree]
It will be difficult to make fine
adjustment when the lock
screw is loosed too much
because the resistance from
the base is lost then.