EasyManua.ls Logo

Hoffrichter LAVI - Page 71

Hoffrichter LAVI
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 6: Alarms and messages 71
Priority Audible
alarm
Info LED
status
Cause Correction
No sys-
tem flow
HIGH
C c c – C c
Red
flashing
No leak in the
ventilation sys-
tem
Use an exhalation
valve for the ven-
tilation mask.
Check the air out-
let with a CPAP
mask.
- -
Low leak
rate
HIGH
c a f – a f
Red
flashing
Leak rate
< 50% than
necessary leak
rate
Check the exhala-
tion valve for the
ventilation mask.
Check the air out-
let with a CPAP
mask.
- -
Motor
speed
low
HIGH
C c c – C c
Red
flashing
Minimum
motor speed
exceeded
Switch device on
and off. If the
error occurs again,
return the device
for servicing
- -
Stenosis HIGH
c a f – a f
Red
flashing
No volume
detected for 3
breaths
Check tube circuit
and tubes for wear
- -
Ventila-
tion off
HIGH
C c c – C c
Red
flashing
Ventilation has
been switched
off due to one
of the following
alarms:
None
System flow
Low leak rate
Motor speed
low
Overcurrent
Overpressure
-
- -
Overpres-
sure
HIGH
c a f – a f
Red
flashing
Pressure switch
triggered five
times within
2s
Restart device. If
the error occurs
again, return the
device for servic-
ing
- -
Overcur-
rent fuse
HIGH
c a f – a f
Red
flashing
Excessive
motor current
Restart device. If
the error occurs
again, return the
device for servic-
ing
- -

Table of Contents