EasyManua.ls Logo

Hoffrichter LAVI - Page 72

Hoffrichter LAVI
102 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72 Chapter 6: Alarms and messages
Priority Audible
alarm
Info LED
status
Cause Correction
Mini-
mum vol-
ume not
reached
HIGH
c a f – a f
Red
flashing
Measured min-
imum volume
less than set
“Minimum Vol-
ume”
-
- -
Internal
commu-
nication
error
HIGH
c c c – c c
Red
flashing
Communica-
tion to control-
ler interrupted
for more than
10 s
Switch device on
and off. If the
error occurs again,
return the device
for servicing
- -
Discon-
nection
HIGH
c a f – a f
Red
flashing
Tube circuit
not connected
to device
Connect the
tube circuit to
the device
- -
Internal
battery
error
MEDIUM c c c Yellow
flashing
Defective bat-
tery
Replace battery
-
-
Internal
Battery
Empty
MEDIUM C c c Yellow
flashing
Device operat-
ing on battery
power supply,
battery charge
≤ 5%
Battery must be
charged; only
5 minutes until
power supply fails
completely; venti-
lation only possi-
ble with external
power supply
- - -
Power
failure
MEDIUM C c c Yellow
flashing
Power supply
from the mains
(AC) connec-
tion has failed
Restore the power
supply
- -
FiO
2
sen-
sor
1
error
MEDIUM C b a Yellow
flashing
FiO
2
sensor
has been sep-
arated or used
up by device
Connect the FiO
2
sensor to the
device
-
-
Low
Internal
Battery
LOW e C Glows
steadily
in tur-
quoise
Device in bat-
tery operation,
battery charge
approx. 20%
2
Charge battery
- -
Humidifier
defective
LOW e C Glows
steadily
in tur-
quoise
Humidifier
defective
Humidifier must
be serviced
- -

Table of Contents