8.4 Periodical inspection
Periodical inspection should be performed to ensure the detection capability of the UST. Table 8-4 below
shows an example list of periodical inspection items. It is recommended to perform this inspection in six
months interval. This inspection should be performed together with daily inspection.
Table 8-4 Periodical inspection list
No.
Check item
Condition
Remark
Yes
No
1
Screw are tightly fastened
2
No displacement from the original mounting position
3
No scratch or crack on the optical window
4
Screws of the optical window are tightly fastened
5
No oil/grease or dirt on the optical window
6
No visible damage on the UST
7
Cable connector is in good condition and tightly fastened
8
When the test piece is placed in the detection area
Output 1 to 3 switch from ON state to OFF state
9
When the test piece is removed from the detection area
Output 1 to 3 switch from OFF state to ON state
10
Area switching is according to input signal
11
Predetermined system operation is carried out as expected
* Must check items 8 and 9 for all the areas
For the traceability purposes it is recommended to maintain and store the