GasAlertMicro 5/PID/IR
Activating the Detector
If any sensor is over due for calibration, the LCD displays the
name of the sensor and the number of days past due.
Due-Lock Enabled
The Due-lock option ensures that a passcode must be entered when
calibration is past due, otherwise the detector automatically deactivates.
11. If Due-Lock is enabled in the user options, the following
screen displays.
Enter the correct passcode and press C to confirm.
Note
If any sensor is overdue, Y displays continually until calibration
is performed.
If no passcode is entered, or it is entered incorrectly, the follow-
ing screen displays.
To enable/disable this option, refer to Due-Lock
in the user
options menu. Also refer to Passcode Protect
.
Force Calibration Enabled
12. If Force cal (force calibration) is enabled in tech mode, calibra-
tion is mandatory before the detector enters normal operation.
Refer to Force Calibration
in Tech Mode to enable/disable, and
refer to Calibration Procedure
to calibrate.
If C is not pressed to start calibration, the following screen dis-
plays and the detector deactivates.