While the detector is spanning the sensor(s), a countdown of
time remaining displays in the lower left of the screen.
When the span is complete, the following screen
displays.
Successful Span: Proceed to Successful Span
step #6.
Unsuccessful Span: Refer to
Unsuccessful Span for possi-
ble solutions.
Successful Span
6. If the sensor(s) has spanned successfully, the audible alarm
beeps three times and the following screens display.
If there are more sensors to span, remove the current cylinder
and connect the next gas cylinder.
Press G and apply gas to span the other sensor(s).
Or
Press H to end the span and proceed to Setting the Calibra-
tion Due Date step #7.
When all sensors have spanned, the following screen displays.
If a sensor fails to span, refer to
Unsuccessful Span at the end of the
calibration section for possible solutions.