GasAlertMicro 5/PID/IR
User Options Menu
Safe Display Sensor Configuration
The Sensor option provides access to additional options and functions
for each sensor.
Depending upon the sensor that is selected, some or all of the following
options are available for configuration:
Enabling/disabling a sensor
Setting the span gas value
Adjusting the STEL period (not applicable to LEL and O
2
sensors)
Selecting the TWA method (not applicable to LEL and O
2
sensors)
Resolution setting (not applicable to CO, LEL, O
2
, and CO
2
sensors)
•% vol CO
2
(CO
2
sensor only)
•% vol CH
4
(LEL sensor only)
Selecting the correction factor (LEL and PID sensors only)
Automatic calibration (O
2
sensor only)
When enabled, the Safe option confirms that
conditions are normal and there are no gas
hazards present. When all gas levels are
normal or below the alarm setpoints, Safe
displays continually on the LCD.
Safe does not display if any active sensor has
failed, if any active sensor has an overdue cal-
ibration, and/or if any active sensor has failed
a bump test.
The detector is shipped with the Safe option
disabled.